<link rel="stylesheet" href="assets/fonts/roboto/roboto.css">
Menu

Fluke 7526A — Precision bench process calibrator

Fluke 7526A combines a precision signal source and an isolated measurement channel for laboratory and workshop calibration. It sources and measures DC voltage, current, resistance, RTD and thermocouple signals; two displays show source and measured results simultaneously.
  • DC voltage output: 0…100 V
  • DC voltage input: 0…100 V; isolated
  • DC current output: 0…100 mA; 1 µA
  • DC current input: 0…50 mA; 0.1 µA; isolated
  • Loop power: 24 V DC ±10%; 0…24 mA
  • Availability: On order
  • Manufacturer: Fluke
  • SKU: 01175
Price: on request

Description

Fluke 7526A combines a precision signal source and an isolated measurement channel for laboratory and workshop calibration. It sources and measures DC voltage, current, resistance, RTD and thermocouple signals; two displays show source and measured results simultaneously. It supports nine RTD types, thirteen thermocouple types, nine stored setpoints per parameter and switch testing. Pressure checks require a compatible external 700P or 750P module and a separate pressure source. RS-232 and IEEE-488 interfaces support automation. Select MET/CAL software and pressure modules separately.

Specifications

DC voltage output
0…100 V
DC voltage input
0…100 V; isolated
DC current output
0…100 mA; 1 µA
DC current input
0…50 mA; 0.1 µA; isolated
Loop power
24 V DC ±10%; 0…24 mA
HART resistor
250 Ω ±3%
Resistance output
5…4000 Ω
Resistance input
0…4000 Ω
Thermocouple mV range
−10…+75 mV; 1 µV
RTD types
9
RTD sensor types
Pt100 (385/3926/3916), Pt200, Pt500, Pt1000, Ni120, Cu427, SPRT
Thermocouple types
B, C, E, J, K, L, N, R, S, T, U, XK, BP
Pt100 α385 simulation
−200…+800 °C
Stored setpoints
9 for each input/output parameter
Display and channels
2 LCDs; simultaneous sourcing and isolated measurement
Pressure function
With an external compatible 700P/750P module and pressure source
Interfaces
RS-232; IEEE-488 (GPIB)
Automation
MET/CAL compatible; software separately
Mains supply
Selectable 100…120 V or 220…240 V; 47…63 Hz; ≤15 VA
Operating temperature
0…+50 °C
Dimensions and weight
146 × 445 × 298 mm; 4.24 kg without options
Standard package
Calibration report, power cord, TC shorting jumper, USB/serial adapter, documentation