Fluke 2271A — Modular industrial pressure calibrator

Fluke 2271A automates calibration of pressure gauges, transmitters and switches. Its modular chassis accepts two PM200 or PM500 measurement modules at the same time. The family spans −100 kPa to 20 MPa; the installed modules determine the actual range.
  • Pressure range: −100 kPa…20 MPa depending on modules
  • Module capacity: 2 × PM200 / PM500
  • PM200 uncertainty: Most gauge ranges: 0.02% FS; low and absolute ranges have separate specifications
  • PM500 uncertainty: For most ranges at 50–100% of range: 0.01% of reading; see module specification
  • Test ports: 2; HC20 or P3000 configuration
  • Availability: On order
  • Manufacturer: Fluke
  • SKU: 01163
Price: on request

Description

Fluke 2271A automates calibration of pressure gauges, transmitters and switches. Its modular chassis accepts two PM200 or PM500 measurement modules at the same time. The family spans −100 kPa to 20 MPa; the installed modules determine the actual range. The built-in EMM electrical module supports HART, 24 V DC loop power, mA and DC voltage measurements. Two test ports and an atmospheric reference port support device connections. Pressure ranges, uncertainty, NPT/BSP manifold and HC20/P3000 ports are selected for the ordered configuration. The CPS contamination prevention system and COMPASS software are options. A suitable clean, dry air or nitrogen supply is required.

Specifications

Pressure range
−100 kPa…20 MPa depending on modules
Module capacity
2 × PM200 / PM500
PM200 uncertainty
Most gauge ranges: 0.02% FS; low and absolute ranges have separate specifications
PM500 uncertainty
For most ranges at 50–100% of range: 0.01% of reading; see module specification
Test ports
2; HC20 or P3000 configuration
Manifold options
NPT / BSP
Reference port
Atmospheric reference
Electrical measurement module
EMM; mA / V DC; HART
Loop power
24 V DC
HART resistor
Built-in switchable 250 Ω
Remote interfaces
Ethernet, RS232, USB
Supply medium
Clean dry air or nitrogen ≥99.5%
Operating temperature
15…35 °C
Warm-up time
Typically 15 min
Ingress protection
IP20
Chassis weight
15 kg
Chassis dimensions
442 × 446 mm; height HC20: 305 mm / P3K: 237 mm
Configuration requirement
Chassis and at least one pressure module; modules selected with the order
Options
CPS contamination prevention; COMPASS software